What You See on the Display
What You Do
Comments
4-3
Questions? Refer to Page 10-10 for Technical Support Instructions.
5.
Press
Scan
to scan the test strip bar
code or manually enter the test strip lot
number via the keypad, then press
Enter.
Scanning the bar code identifies the strip
type, calibrates the monitor, ensures the
expiry date has not passed, records the strip
lot used, and checks that the lot has been
approved for use by your institution.
6.
Open the foil test strip packet at the
notch and tear down to remove the test
strip.
With clean, dry hands, you may gently touch
the test strip anywhere when removing it
from the foil to insert it into the monitor.
7.
With the contact bars facing up, insert the
test strip into the test strip port until it
stops and
Strip Inserted
is displayed.
Prior to inserting the test strip, ensure that
the port protector is installed and that it
is clean and dry. Replace if necessary. The
display will alternate between
Insert Strip
and
Glucose.
8.
Gently invert the required control
solution bottle 3-4 times. Remove the cap
of the control solution bottle and wipe the
nozzle with a clean gauze or tissue. Apply
a small drop of solution to the test strip
target area, allowing the target area to fill
completely. Wipe the nozzle of the control
solution bottle before replacing the cap.
When sufficient sample has been applied,
the monitor beeps, displays
Sample
Accepted
and automatically starts the test.
For a detailed description of the test strip
target area, refer to the Test Strips section of
Chapter 1.
/
Note:
Place monitor on a flat surface while running control tests.
Note:
Do not lift the monitor to view the display until after the test strip has been removed. Doing so may cause
control solution to drip onto the port protector or into the strip port, damaging the monitor.
If the port protector becomes wet, replace it with a new one.
Chapter 4 : Control Test
Содержание Precision Xceed Pro
Страница 37: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test β Ketone ...
Страница 42: ...7 Linearity Test ...
Страница 60: ...9 Data Upload Docking Station Mounting Instructions Enabling Data Upload ...
Страница 77: ...11 Maintenance Cleaning the Exterior Surface Replacing the Port Protector Battery Life Replacing the Batteries ...
Страница 80: ...12 Specifications Precision Xceed Pro Monitor Symbols Bar Code Types Important Safety Instructions ...