1-5
Questions? Refer to Page 10-10 for Technical Support Instructions.
Calibration Verification Controls
Use RNA Medical brand Glucose and β-Ketone Calibration Verification Controls to confirm the calibration and analytical
measurement range of the monitor for Precision Xceed Pro Blood Glucose or Blood β-Ketone Test Strips. Use RNA Medical
brand Glucose Calibration Verification Controls with Precision PCx Plus Blood Glucose Test Strips. This is called linearity
testing and requires a kit with 5 levels of control solution. Linearity testing is further described in Chapter 7.
Isolation Bag (optional)
Abbott Isolation Bags are disposable plastic bags for use with the Precision Xceed Pro System. The Isolation Bag provides
a partial barrier between the point-of-care monitoring device and the patients in isolation for infection control. There is
no need to prep the monitor prior to using the bag and the bag does not affect any monitor functions. Note that Abbott
Isolation Bags are not sterilized.
Carry Case (optional)
The carry case holds the monitor, test strips, control solutions, the
quick reference guide and other accessories needed for
testing. A transparent cover allows you to check supply
status without opening the case.
Docking Station (optional)
The docking station provides a convenient way
to upload the data from the monitor to the data management
system. When the monitor is placed in the docking station, it will automatically upload
data to the data management system and then shut down. Data upload is further
described in Chapter 9.
Data Upload Cable (optional)
Alternatively, a data upload cable is available to connect the monitor to the computer
running the data management system.
Data Management System
A computer running a data management system is necessary to configure and upload
results from the Precision Xceed Pro Monitor. Abbott’s PrecisionWeb Point of Care Data Management System and software
from other vendors are available. Data upload from remote locations requires either Terminal Servers or networked PCs
running Abbott's Data Repeater software.
Chapter 1: Overview
Содержание Precision Xceed Pro
Страница 37: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test β Ketone ...
Страница 42: ...7 Linearity Test ...
Страница 60: ...9 Data Upload Docking Station Mounting Instructions Enabling Data Upload ...
Страница 77: ...11 Maintenance Cleaning the Exterior Surface Replacing the Port Protector Battery Life Replacing the Batteries ...
Страница 80: ...12 Specifications Precision Xceed Pro Monitor Symbols Bar Code Types Important Safety Instructions ...