Questions? Refer to Page 10-10 for Technical Support Instructions.
Battery Life
The average alkaline battery life is approximately 60 days, based on an average of 9 tests per day and monitor being shut
off within 2 minutes of it not being used (depending on bar code usage). Battery life will vary based upon actual use.
Replacing the Batteries
Use the following procedure to install new batteries in the Precision Xceed Pro Monitor. The Precision Xceed Pro
Monitor requires two (2) AA alkaline batteries or nickel-metal hydride (NiMH) rechargeable batteries (not supplied by
Abbott Diabetes Care).
1. Turn the Precision Xceed Pro Monitor off by pressing the On/Off
button on the keypad.
2. Turn the monitor over and remove the battery door.
3. Pull the blue tab to remove the batteries (Figure 1).
4. Discard the used batteries in compliance with your local government regulations.
5. Insert the new batteries on top of the strap, using the + and - symbols in the battery compartment to position the new
batteries with the correct polarity. Refer to Figure 2 below.
6. Align the battery compartment cover with the slots on the Precision Xceed Pro Monitor, then snap the cover into
place.
Figure 1
Figure 2
IMPORTANT
: Rechargeable batteries must be at room temperature before they are placed in the battery
compartment. The docking station does not charge the batteries.
Chapter 11: Maintenance of the Precision Xceed Pro Monitor
Replacing the Port Protector
The port protector is designed to minimise the possibility of liquid entering the monitor
through the strip port. Should blood or control solution come into contact with the port
protector, the monitor should be cleaned and dried and the port protector replaced. Be
sure to dry the area around the test strip port prior to installing the new port protector.
Removal
1. Place the monitor on the flat surface during the removal process so that the port
protector is in full sight. Place one hand at the bottom of the monitor and with the other
hand insert a common flat tool device, such as a small screw driver, under the left or
right edge of the port protector. Do not insert the flat tool device into the strip port
opening. A standard flat tool device is recommended (not included).
(Fig. 1)
2. Carefully prise the port protector until the protector separates from the monitor.
3. Pull the port protector away from the monitor and discard as per your facility’s
biohazard disposal policy.
(Fig. 2)
Insert
1. Obtain a new port protector.
2. Rest the flat bottom of the new port protector on the ledge of the test strip port.
(Fig. 3)
3. Push both sides of the port protector until the tabs snap into place. Note that some force
is required to fix the port protector onto the monitor.
(Fig. 4)
1
2
3
4
Содержание Precision Xceed Pro
Страница 37: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test β Ketone ...
Страница 42: ...7 Linearity Test ...
Страница 60: ...9 Data Upload Docking Station Mounting Instructions Enabling Data Upload ...
Страница 77: ...11 Maintenance Cleaning the Exterior Surface Replacing the Port Protector Battery Life Replacing the Batteries ...
Страница 80: ...12 Specifications Precision Xceed Pro Monitor Symbols Bar Code Types Important Safety Instructions ...