What You See on the Display
What You Do
Comments
7-4
Questions? Refer to Page 10-10 for Technical Support Instructions.
12.
Follow the instructions in the CVC kit
package insert. Then apply a drop of the
sample to the target area.
When sufficient sample has been applied,
the monitor beeps, displays
Sample
Accepted
and automatically starts the test.
For a detailed description of the test strip
target area, refer to the Test Strips section of
Chapter 1.
13.
Wait for the monitor to analyze the
sample and display the test result.
The monitor counts down then displays
the test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
14.
Note the result. If required, scan or
manually enter the comment code, and
press
Enter.
If there is no prompt to enter a comment
code, skip down to step 15.
The monitor may be enabled to scan or
enter a 1- to 2-digit comment code.
15.
You can select one of the following
options:
Press
1
–
New Level
.
Press
2
–
Same Level
.
Press the
Menu
button to return to the
Menu Mode
.
Press
On/Off
to turn off monitor.
Remove the test strip from the monitor
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
The display shows:
Level
(test level 1 through 5)
Test
(replicate 1 through 4 for that level)
Comment Code
(Number in box)
Date and Time of test
Operator ID
(
o:
prompt)
Note that test date and time alternates with
Operator ID.
If
2
is selected, the
Scan or Enter Strip
Lot
screen will appear. Return to step 8 to
continue linearity testing with the
Same
Level
.
If
1
is selected, the
Select Level
screen will
appear. Return to step 7 to perform linearity
tests for a
New Level
.
45 mg/dL
45 mg/dL
Chapter 7: Linearity Test
Содержание Precision Xceed Pro
Страница 37: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test β Ketone ...
Страница 42: ...7 Linearity Test ...
Страница 60: ...9 Data Upload Docking Station Mounting Instructions Enabling Data Upload ...
Страница 77: ...11 Maintenance Cleaning the Exterior Surface Replacing the Port Protector Battery Life Replacing the Batteries ...