Pannoramic 250 FLASH III 3.0.2 User’s Guide
5.1.1 Profile Settings Panel
Figure 28 – Preview image with and without coverslip edge
Caution!
It requires lots of time (since it cannot find focus neither on the edges,
nor on specks) to digitize coverslip edges. Select
Remove coverslip
function to eliminate those areas.
Warning
!
Do
NOT
use
Remove coverslip
function on TMA slides! The layout of
TMA slides may mislead the detecting algorithm.
◦
Fill holes inside the scanmap
–
If selected, the holes inside the specimen will also be
scanned. Often these holes contain small sample objects that otherwise would not be
scanned.
◦
Remove specks
–
If selected, the scanner does not digitize objects smaller than the
predetermined size. This helps filtering specks and other uninteresting objects out. Set
the maximum size of specks not to be digitized by adjusting the slider.
◦
Dilate mask
–
Adjust the slider to enlarge or decrease the scanned area. The edge is
extended discretely by 100 µm each time.
Scanmap
A previously saved scanmap can be loaded and applied.
•
Use Threshold:
Preview settings defined in the Microscope Controller Panel / Preview
window are applied as the basis of scanmap.
September 16, 2020 – Rev. 1
3DHISTECH Ltd.
60(156)