Pannoramic 250 FLASH III 3.0.2 User’s Guide
1 Product Description
1 Product Description
1.1 Product Overview
1.1.1 Intended Use
Pannoramic® 250 FLASH III is designed to use transmitted light and reflected light fluorescence to
digitize biological samples and specimens (typically histological sections). The target users of the
product can work for example, in education, research, industry or medical sectors.
Warning!
After each scanning make sure to inspect the digital slide for over-exposed
areas. If the slide is scanned based on improperly set parameters, it can
highly affect the diagnostic result because of information loss.
Warning!
Any modification to a hardware or software component of the PC will highly
affect the functionality of the scanner program.
1.1.2 Features and Benefits
The Pannoramic® 250 FLASH III offers world class brightfield and fluorescent whole slide scanning on
up to 9 channels. An outstanding 0.19 µm/pixel (in Brightfield) and 0.16 µm/pixel (in Fluorescent)
resolution is achieved with a 40×/0.95 NA (equivalent to 52× magnification in Brightfield and 62×
magnification in Fluorescent modes) Plan-Apochromat objective.
The key features of the Pannoramic® 250 FLASH III Scanner are as follows:
•
Extreme fast brightfield scanning mode:
•
Continuously moving stage during scanning
•
Brightfield camera with a frame rate of up to 180 fps
•
Xenon flash lamp as light source
September 16, 2020 – Rev. 1
3DHISTECH Ltd.
14(156)