Pannoramic 250 FLASH III 3.0.2 User’s Guide
7 Troubleshooting and Maintenance
5. Hold the filter module
(3)
by its handling pads (one on the right, one on the left (lateral)
side of the module) and insert it in forward tilted position from above into the upper spring-
loaded clamps
(1)
at the filter wheel.
Warning!
Avoid touching the surface of the filters, and store the modules in their
own boxes when not used.
6. Press the filter module down, until it safely connects to the bottom spring-loaded clamps
(2)
of the filter wheel. Mounting position of the filters: excitation filter
(5)
and emission filter
(4)
7. To remove the filter module, first slightly tilt it out of its upper spring-loaded fixtures, then
pull it out of the bottom spring-loaded components and take it off.
8. Close the protective cover (see chapter
7.4.1 Opening the protective cover
).
9. Power on the scanner. See chapter
2.4 Power the scanner on/off
on how to power on the
device.
10. Enter the respective parameters in the input fields of the software (see chapter
).
11. Turn the fluorescent illumination unit on again.
7.4.5 Removing a Magazine from the Transporting System
Figure 108 – Removing a magazine
1. Open the protective cover (see chapter
7.4.1 Opening the protective cover
).
September 16, 2020 – Rev. 1
3DHISTECH Ltd.
147
(