Pannoramic 250 FLASH III 3.0.2 User’s Guide
3.3.2 Insert and remove magazines
Warning!
The protective cover must always be closed while the microscope is
loading and/or scanning!
Note:
The guide rails have built-in sensors that detect whether there are
magazines inserted in the input stack, or whether the output stack is
full. If the system does not detect the inserted magazine in the input
stack, check whether you have inserted it correctly. If the magazine is
placed correctly, there may be a problem with the sensor. In this case,
contact 3DHISTECH Service Support.
1. Remove magazine cover, before you insert magazines.
2. Put a magazine
(2)
into the input stack
(3)
. The magazine’s rack (the back side of the
magazine)
(1)
must contact with the guide rails
(5)
, with its open side towards the scanner.
NOTE:
The
symbol at the magazine feeder refers to the possibility of exposure to
biological hazards whenever handling slides.
3. Slide down the magazine on the input guide rails until it rests on the gear wheels
(6)
or
magazines.
Figure 15 – Gear wheels
September 16, 2020 – Rev. 1
3DHISTECH Ltd.
35
(
156
)