3-12
SETTING-UP TESTS
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
6
Repeat Steps 4-5 for the
left eye.
Note: SPHERE, CYLINDER, and AXIS may be chosen in any sequence. To correct entries, re-
select the command button and then enter the correct data.
7
Select CALCULATE TRIAL LENS.
8
The trial lens data is
automatically entered on the
Patient Data 1 screen.
FOR MANUAL TRIAL LENS INPUT:
1.
From the Patient Data 1 screen, select TRIAL LENS.
2.
Choose MANUAL TRIAL LENS INPUT.
3.
Repeat Steps 3-6 above.
4.
Choose ENTRY COMPLETE. The manually entered trial lens data is automatically entered on
Patient Data 1.
Note: For guidelines on selecting the proper trial lens (for manual input), refer to “Using
Trial Lenses” and Table 3.4.