11-6
KINETIC TESTING
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
Note: Asterisks (*) below denote the default setting.
Stimulus Size
This drop-down menu allows the user to set the size of the test stimulus. Sizes can be set to one
of the five standard Goldmann test stimulus sizes. They are designated by Roman numerals I*
through V.
Stimulus Intensity
This drop-down menu lists numbers corresponding to the first (“coarse”) Goldmann intensity
filter, which helps to determine the intensity of the stimulus. The values corresponding to the
various standard Goldmann filters are:
Goldmann
Humphrey
Filter Number
Intensity
1
25 dB
2*
20 dB
3
15 dB
4
10 dB
Stimulus Intensity Modifier
This drop-down menu lists the letters corresponding to the second (“fine”) Goldmann
intensity filter, which also helps to determine the intensity of the stimulus. The values
corresponding to these filters are:
Goldmann
Humphrey
Goldmann
Humphrey
Filter Letter
Intensity
Filter Letter
Intensity
A
+4 dB
A´
+24 dB
B
+3 dB
B´
+23 dB
C
+2 dB
C´
+22 dB
D
+1 dB
D´
+21 dB
E*
+0 dB
E´
+20 dB
The actual intensity of the test stimulus is equal to the combined values of the Stimulus
Intensity and the Stimulus Intensity Modifier. For example, a stimulus with an Intensity of “3”
and Intensity Modifier of “C” will have a resultant intensity of 17 dB (15 + 2).
Stimulus Speed
This drop-down menu lets the user set the speed, in degrees per second, at which the stimulus
moves in the test bowl. Users can choose any test speed between 1 and 9 degrees per second.
11.htp
3/13/03, 3:00 PM
6
Summary of Contents for HUMPHREY 720i
Page 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Page 4: ......
Page 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 327: ......
Page 328: ......