HUMPHREY FIELD ANALYZER
II -
i
series
HUMPHREY FIELD ANALYZER
II
APPENDIX H
H-3
PROBLEM
POSSIBLE CAUSE OR SOLUTION
Printer
No signal from HFA II
• Check printer cable connection (1)
• Check cable is correct and not damaged
• Check to ensure correct printer is selected on System Setup menu (2)
• Printer buffer full - turn off printer and turn back on
• No printout selected to print (7)
• Other printers: Consult specific printer manual
Improper printout
• Random characters or multiple pages printed
- Check that correct printer is selected on System Setup menu (2)
- Printer needs to be turned off to be reset
• Abnormal value printed - reprint visual field
(printer occasionally prints double characters)
• Printrex printout blank - paper may be upside-down
Printouts fade over time
• Do not store printouts in plastic covers or use adhesive tape on
thermal paper (12)
Printout Problems
No change probability symbols on
• GCP requires 3 tests to show change probability symbols (7)
Glaucoma Change Probability (GCP)
printout
No Single Field Analysis printout
• Test pattern not eligible for STATPAC analysis (7)
• One or more test parameters not valid for STATPAC (7)
Quad totals not shown
• Available on Three-in-One printout only (7)
No SF or CPSD values on
• SF and CPSD are not displayed when SITA test strategy is used (4, 7)
threshold printout
• Fluctuation was turned off for the test (CPSD is not calculated when
fluctuation is off)
Glaucoma Hemifield Test (GHT)
• FastPac test strategy used (GHT not available with FastPac) (7)
results not displayed
• Test not eligible for STATPAC analysis (7)
Printout does not include all eligible
• Some test parameters not valid for STATPAC (7)
tests for a patient when printing
• Patient name or date of birth incorrect on some tests (3, 8)
Overview or Glaucoma Change
• Search via RECALL PATIENT DATA - fix with Merge Patients feature (3, 8)
Probability (GCP) printout
• Sixteen (16) test maximum for Overview and GCP printouts (7)
• FastPac, SITA Standard and SITA Fast tests are ineligible for Glaucoma Change
Probability Analysis (7)
B-H Appendex.htp
3/14/03, 3:24 PM
27
Summary of Contents for HUMPHREY 720i
Page 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Page 4: ......
Page 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 327: ......
Page 328: ......