11-14
KINETIC TESTING
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
PRE-DEFINED
KINETIC TEST
PATTERNS
Your Humphrey Field Analyzer is programmed with a number of pre-set kinetic test patterns.
The following 3 patterns will automatically test the meridians listed below. Only one isopter
(III 4 E) is available with these set patterns. Special mapping options are only available when
the test is run in the Step by Step mode.
Standard 45 (45 degree steps & 5 degrees on either side of midlines)
Speed: 2 degrees per second
Size: Goldmann III4e
Total Number of Points per Isopter: 12
Meridians Tested : 5
°
, 45
°
, 85
°
, 95
°
, 135
°
, 175
°
, 185
°
,
225
°
, 265
°
, 275
°
, 315
°
, 355
°
Standard 30 (30 degree steps & 5 degrees on either side of midlines)
Speed: 2 degrees per second;
Size: Goldmann III4e
Total Number of Points per Isopter: 15
Meridians Tested: 5
°
, 30
°
, 60
°
, 85
°
, 95
°
, 120
°
, 150
°
, 175
°
,
185
°
, 210
°
, 240
°
, 265
°
, 275
°
, 300
°
, 330
°
High Res 15 (15 degree steps & 5 degrees on either side of midlines)
Speed: 2 degrees per second;
Size: Goldmann III4e
Total Number of Points per Isopter: 28
Meridians Tested: 5
°
, 15
°
, 30
°
, 45
°
, 60
°
, 75
°
, 85
°
, 95
°
, 105
°
,
120
°
, 135
°
, 150
°
, 165
°
, 175
°
, 185
°
, 195
°
,
210
°
, 225
°
, 240
°
, 255
°
, 265
°
, 275
°
, 285
°
,
300
°
, 315
°
, 330
°
, 345
°
, 355
°
In addition to the 3 patterns listed above, there are two other special Kinetic tests available.
One is the Social Security Administration Disability Test (SSA Test). The other is a special test
used in Germany (BG/FS Test). The BG/FS test is identical to the Standard 30 test listed above.
Press the desired test button
found in the Kinetic Tests library.
You will have the option to run
the test in the Automatic or the
Step by Step mode as described
next.
11.htp
3/13/03, 3:01 PM
14
Summary of Contents for HUMPHREY 720i
Page 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Page 4: ......
Page 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 327: ......
Page 328: ......