STATPAC ANALYSIS & PRINTING RESULTS
7-3
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
STATPAC test
STATPAC will analyze tests that fall within the parameters listed below:
parameters
Table 7.1: STATPAC Parameters for White-on-White Perimetry
Type of test:
Threshold
Test pattern:
Central 10-2, 24-2, 30-2
Test strategy:
SITA Standard, SITA Fast, Full Threshold, FastPac
Stimulus color:
White
Stimulus size:
Size III
Fixation target:
Any
Foveal threshold:
On or Off
Fluctuation test:
On or Off (SITA tests automatically set to Off)
Test Speed:
Normal or Slow
STATPAC analysis may be used with all Central 24-2 and 30-2 threshold test results. There are
some limitations. The Glaucoma Change Probability Analysis and Glaucoma Hemifield Test
(GHT) are not available with tests using the FastPac strategy. Glaucoma Change Probability
Analysis is also not available for the SITA tests. For Central 10-2 test results, STATPAC produces
a Single Field Analysis or an Overview showing up to sixteen (16) tests results; the Change
Analysis and Glaucoma Change Probability Analysis are not available.
The parameters needed for STATPAC analysis of Blue-Yellow test results are listed below. Single
Field Analysis and Overview printouts are available. The GHT is not available with FastPac tests.
Table 7.2: STATPAC Parameters for Blue-Yellow Perimetry
Type of test:
Threshold
Test pattern:
Central 24-2, 30-2
Test strategy:
Full Threshold, FastPac
Stimulus color:
Blue
Stimulus size:
Size V
Fixation target:
Any
Foveal threshold:
On or Off
Fluctuation test:
On or Off
Test Speed:
Normal or Slow
7.htp
3/14/03, 3:17 PM
3
Summary of Contents for HUMPHREY 720i
Page 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Page 4: ......
Page 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 327: ......
Page 328: ......