GENERAL OPERATION
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
2-15
Accessing the save/transmit
1
Start at the System Setup screen. Select the SAVE / TRANSMIT OPTION. Press the TRANSMIT
option - serial connection
VIA button to toggle between Serial and Network connection settings. Network
connections are described on the next page.
2
The drop-down menu at
the right allows you to
choose from SAVE ONLY, SAVE
AND TRANSMIT, or TRANSMIT
ONLY.
3
After choosing SAVE AND
TRANSMIT or TRANSMIT
ONLY, press SERIAL TRANSMIT
SETTINGS. The Transmit RS-232
Setup screen will appear.
4
At the Transmit RS-232
screen select the Baud Rate,
Parity, Data Bits and Stop Bits
as required for transmission.
The following choices are available:
Baud Rate
300, 600, 1200, 2400, 4800,
9600
, 19200
Parity
none
odd
even
Data Bits
7
8
Stop Bits
1
2
For compatibility with other Carl Zeiss Meditec products (Ensemble, HFA 600 series instru-
ments), transmit serial data using 9600 Baud at Even parity with 7 data bits and 1 stop bit.
5
Select PROCEED to save the changes or CANCEL to restore the previous values. Press
PROCEED on the first screen to save the settings and return to the System Setup screen.
2.htp
3/14/03, 1:51 PM
15
Summary of Contents for HUMPHREY 720i
Page 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Page 4: ......
Page 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 327: ......
Page 328: ......