GENERAL OPERATION
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
2-13
Accessing head tracking
1
Start at the System Setup screen. Select the Head Tracking drop-down menu.
(model 750i)
2
Choose between ON and
OFF.
Note:
If Head Tracking is turned on during testing and the patient moves, the instrument
will adjust the chin rest in small, 0.3 mm increments until the patient returns to the
original position. This feature only works if Gaze Tracking has been successfully initial-
ized and the trial lens holder is in the Up position. For additional information, see
Section 5: “ Head Tracking”.
Accessing the vertex
1
Start at the System Setup screen. Select the Vertex Monitor drop-down menu.
monitor (model 750i)
2
Choose between ON and
OFF.
Note: When the Vertex Monitor is turned on, a beep will sound if the patient has backed
away from the trial lens during testing. Although the test will not pause, a message will
remain on the screen until cleared by the operator. This feature only works if Gaze
Tracking has been successfully initialized and the trial lens holder is in the Up position.
For additional information, see Section 5: “Vertex Monitor”.
2.htp
3/14/03, 1:51 PM
13
Summary of Contents for HUMPHREY 720i
Page 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Page 4: ......
Page 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 327: ......
Page 328: ......