Appendix D: Goldmann Conversion Tables
R E V . A 10 9/ 9 8 P N 47653
H U M P H R E Y F I E L D A N A L Y Z E R
I I
D-1
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
The decibel notation the Field Analyzer uses can be expressed in terms of Goldmann units or
apostilbs. The conversion between decibels (dB) and apostilbs (asb) and Goldmann units are
listed in Table D-1. For example, a threshold sensitivity of 25 dB using a size III stimulus is equal
to a Goldmann III-Ie stimulus or 32 asb. A threshold sensitivity of 0 dB using a size III stimulus is
equivalent to a Goldmann V 4e stimulus or 10,000 asb. A simplified chart to convert from dB
values to Goldmann units is available in Tables D-1 through D-6.
To use the conversion tables, first locate the table which lists the target size you used for the test.
For example, you may have completed a central 30-2 threshold test using a size III target and you
want to find out what the Goldmann equivalent is for 15 dB. Using Table D-4, find 15 dB. Read
the horizontal and vertical columns that intersect at 15 dB. The Goldmann equivalent is either a II
4e, III 3e, IV 2e, or V 1e.
Stimulus Size
Angular Subtense
Stimulus Area (30 cm Bowl)
(degrees)
(mm
2
)
Size I
0.11
1/4
Size II
0.22
1
Size III
0.43
4
Size IV
0.86
16
Size V
1.72
64
B-H Appendex.htp
3/14/03, 3:22 PM
7
Summary of Contents for HUMPHREY 720i
Page 1: ...HUMPHREY FIELD ANALYZER II i series USER S GUIDE Model 720i Model 740i Model 745i Model 750i...
Page 4: ......
Page 26: ...1 18 INTRODUCTION INSTRUMENT SETUP REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 116: ...6 8 TEST RELIABILITY REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 208: ...10 20 CUSTOM TESTING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 284: ...12 14 CARE AND CLEANING REV B 02 03 PN 51680 1 HUMPHREY FIELD ANALYZER II i series...
Page 327: ......
Page 328: ......