TEST RELIABILITY
6-5
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
Figure 6.1: Sample Printout Showing Poor Reliability.
The example above indicates a very unreliable patient. A high number of fixation losses, false
positive errors, and false negative errors have been recorded. Poor fixation is indicated by the
gaze graph. Also note the number of points where threshold results are 40 dB or higher.
False Positive
False Negative
Fixation Losses
Gaze Graph
XX
Abnormally high results
Carl Zeiss Meditec
6.htp
3/14/03, 3:16 PM
5