9-14
DATABASE MANAGEMENT
REV. B 02/03 PN 51680-1
HUMPHREY FIELD ANALYZER
II -
i
series
2
Read the warning message.
Select YES. Selecting NO
returns you to Step 1.
3
Continue saving test results to floppy disk until
you have the time to engage the recovery
program.
Note: Each time you attempt to save data or try to use a function that involves the hard
drive, the previous two screens will be displayed. Continue to choose FLOPPY ONLY and
YES to save data to the floppy disk while patients are present.
4
When you have time to
address the database
problem, and only after all tests
have been saved to floppy disk,
recycle power on the HFA II.
Recycling power will initiate a
recovery program. If recovery is
successful, a user message
appears. Select OK.
Copy to your hard drive any test results stored to the floppy drive since the hard disk
failure.