Axioskop 2 FS
plus
START-UP
Axioskop 2 FS
MOT
Initial start-up
Carl Zeiss
B 40-076 e 02/01
2-13
Compensation of ametropia when eyepiece reticles are used
The correct use of an eyepiece reticle requires two focusing eyepieces, e.g. E-PL 10x/23 Br. foc., to allow
the user to compensate for differences in the visual performance of his two eyes.
•
Use the focusing eyepiece to focus on the line figure of the eyepiece reticle; focus on the edge of the
field of view if no eyepiece reticle is used.
•
Use the focusing drive to focus on the microscope image of a specimen through the eyepiece set as
described above.
•
Then use the second focusing eyepiece to focus the microscope image for the second eye. The
position of the focusing drive on the microscope stand must not be changed.
2.1.10
Setting of the interpupillary
distance of the binocular tube
The eyepiece distance is matched to the user's
individual interpupillary distance by swinging the
eyepiece tubes symmetrically towards one another
(2-13/
A
and 2-13/
B
).
2.1.11
Setting of viewing height
The viewing height can be set continuously in the
range from 6 to 25° by changing the viewing
angle of the two ergotubes. Furthermore, the
binocular component of the ergotubes can be
swivelled by 180° .
Fig. 2-13
Setting the interpupillary distance
of the binocular tube
Summary of Contents for Axioskop 2 FS plus
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...