Axioskop 2 FS
plus
START-UP
Axioskop 2 FS
MOT
Attachments and conversions
Carl Zeiss
B 40-076 e 02/01
2-25
2.2.3
Changing the DIC prism in the
universal condenser
•
To loosen the cover on the condenser underside
(2-32/
1
), loosen both grub screws using the
SW
2
screwdriver and remove the cover.
Position the turret disk containing the DIC prism
to be exchanged in the exchange opening and
hold it on the knurled ring.
•
Unscrew retainer ring using the so-called
double-function tool (2-32/
4
).
•
Now screw the tool in the prism mount
(2-32/
3
) with the threaded side (2-32/
4
) and
pull out DIC prism (2-32/
2
).
•
Remove DIC prism (2-32/
2
) from the tool and
screw on the new, required DIC prism.
•
Installation of the DIC prism is made in reverse
order. Make sure that the DIC prism is oriented
correctly (groove 2-32/
5
must engage in the
pinion of the mount). Make sure that the
knurled ring of the turret disk is labelled
correctly.
2.2.4
Attachment of filter mount
•
Hold filter mount (2-33/
5
) parallel to the
underside of the condenser carrier (2-33/
3
) and
screw holding pin (2-33/
4
) of the filter mount
(2-33/
5
) into the front threaded hole on the left
below the condenser carrier (2-33/
3
) until stop
using the SW 2 angled Allen key (2-33/
6
).
•
Use adjusting lever (2-33/
1
) to screw stop bolt
(2-33/
2
) into the rear threaded hole of the
condenser carrier until stop.
Fig. 2-32
Changing the DIC prism
Fig. 2-33
Attachment of filter mount
Summary of Contents for Axioskop 2 FS plus
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...