Axioskop 2 FS
plus
INTRODUCTION
Axioskop 2 FS
MOT
Notes on instrument safety
Carl Zeiss
B 40-076 e 02/01
0-9
The Axioskop 2 FS
plus
and Axioskop 2 FS
MOT
microscopes, including original accessories, may only be
used for the microscope techniques described in this manual.
Particular attention must be paid to the following warning notes:
The manufacturer cannot assume any liability for any other applications, possibly also involving
individual modules or single parts. This also applies to all service or repair work which is not
carried out by authorized service personnel. Furthermore, this forfeits all the claims against
warranty.
The Axioskop 2 FS
plus
and Axioskop 2 FS
MOT
microscopes are provided with a special
transport lock at the front of the stand. The microscopes must not be transported or moved
without this transport lock. It is particularly important to avoid holding the instrument on the
objective slider even if it is moved only a small distance away from its position; otherwise, the
instrument will be damaged.
The power plug must be inserted in a socket featuring a grounding (earth) contact. The
grounding effect must not be made ineffective by an extension cable which does not have a
protective ground wire.
If it is established that the protection measures are no longer effective, the instrument must be
switched off and safeguarded against inadvertent operation. Contact a Zeiss service agency or
Carl Zeiss microscopy service for the repair of the instrument.
☞
The separate 12 V DC 100 W power unit for the HAL 100 permits the use of line voltages in
the range between 100 and 240 V ±10 %, 50 - 60 Hz. Voltage change is not required.
The power supply for the HBO 103 (ebq 100 dc) or XBO 75 (ebx 75 isolated) permits a voltage
range of 90 to 250 V AC, 50 to 60 Hz. The instruments will automatically adjust to the existing
voltage. Voltage change is not required.
Before switching on the instrument, check whether it is suitable for the line voltage present.
Always disconnect the instrument from the line before opening the instrument and before
changing the fuses.
Make sure to use only fuses of the rated power required. The use of makeshift fuses and the
short-circuiting of the fuse holders are not permitted.
Summary of Contents for Axioskop 2 FS plus
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...