Axioskop 2 FS
plus
START-UP
Axioskop 2 FS
MOT
Attachments and conversions
Carl Zeiss
B 40-076 e 02/01
2-23
2.2
Attachments and conversions
☞
The steps required for attachments to and conversions of the microscope are explained in the
following. After this, the respective modules must be reset in a functioning status again.
2.2.1
Changing the condenser
It is recommended to remove the stage carrier if
you want to change the condenser.
•
Remove the stage carrier (see 2.1.4) and safely
place it on the table.
•
Use lever (2-30/
8
) to fold out the front lens on
the available condenser.
•
Move condenser carrier (2-30/
2
) to the lowest
position via drive for height adjustment
(2-30/
5
).
•
Unscrew both centering screws (2-30/
3
) until
condenser can be easily taken out of its mount.
•
Lift available condenser from the orientation
groove by slightly tilting it upwards and pull it
out to the front.
•
Use lever (2-30/
8
) to fold out the front lens on
the condenser to be inserted.
•
Insert condenser (2-30/
1
) between condenser carrier (2-30/
2
) and stage carrier (2-30/
4
). Orient the
locking screw on the underside of the condenser in the direction of the groove (2-30/
7
).
•
Press dovetail of condenser against the spring mount (2-30/
6
) of the condenser carrier until the
condenser can be placed on the condenser carrier horizontally.
•
Let condenser go smoothly; the screw will lock in position in the groove (2-30/
7
) at the front.
•
Screw in centering screws until they engage in the dovetail.
•
Attach stage carrier to the microscope stand again (see section 2.1.8).
Fig. 2-30
Changing the condenser
Summary of Contents for Axioskop 2 FS plus
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...