Axioskop 2 FS
plus
START-UP
Axioskop 2 FS
MOT
Attachments and conversions
Carl Zeiss
B 40-076 e 02/01
2-31
•
Press stage to the front against spring pin and lower its back in the stage carrier (2-42/
3
), then let go.
•
Tighten screwing cap (2-42/
2
).
c)
Centering the stage
When objectives with a high magnification are
used, centering can be exact for one selected
objective only. All stages are factory-precentered,
i.e. a set specimen detail remains in the image
center even when the stage is rotated. If the
image detail moves from the image center (2-43/
5
)
after stage rotation, recentering should be
performed as follows:
•
Loosen stage clamping screw (2-43/
4
) and
screwing cap of the stage carrier (2-43/
1
)
•
Rotate the stage to determine the maximum
specimen deflection (2-43/
5
, tip of the arrow)
in the direction of the eyepiece reticle.
•
Reset the two centering screws on the stage
carrier (2-43/
2
) using one SW 1.5 Allen
screwdriver (2-43/
3
) each to move the
specimen detail by half the arrow length in the
direction of the crossline center. Check whether
specimen detail moves when the stage is
rotated again; repeat the procedure, if
required.
•
When centering is finished, tighten screwing
cap (2-43/
1
) again.
The stage can be rotated by 240° to a range of y
≤
27 mm. No rotation is possible above this range.
Fig. 2-43
Centering the rotary mechanical
stage
Summary of Contents for Axioskop 2 FS plus
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...