Axioskop 2 FS
plus
OPERATION
Axioskop 2 FS
MOT
Operation and function controls
Carl Zeiss
B 40-076 e 02/01
3-7
Lever to swing in / out the diffusion disk
The integrated diffusion disk (3-1/
12
) ) is used to achieve homogeneous illumination, though at a slight
loss of light.
−
Lever in front position:
diffusion disk swung out
−
Lever in rear position:
diffusion disk swung in
Drive for movement of the mechanical stage in Y
−
Movement of the mechanical stage in Y, maximum travel range: 50 mm
−
Adjustable torque (smoothness)
Drive for movement of the mechanical stage in X
−
Movement of the mechanical stage in X, maximum travel range: 75 mm
−
Adjustable torque (smoothness)
3 levers to swing in / out the filters in the transmitted-light filter magazine
−
3 switchable neutral-density filters to match the image brightness in the field of view,
maximum attenuation factor: 1:4400;
Lever correspond to (on the stand from front to back): 1.5 %, 6 % and 25 % transmission
−
lever in front position:
filter swung out
−
lever in rear position:
filter swung in
Centering screw for condenser
−
Centering screws for condenser centering attached to both sides of the condenser carrier.
Universal condenser
Depending on the version, the achromatic-
aplanatic universal condenser (3-2/
1
) is equipped
with:
−
Fold-out front lens
−
Turret disk for:
brightfield - without or with DIC
I
,
II
,
III
darkfield -
D
phase contrast - Ph
1
, Ph
2
, Ph
3
interference contrast - DIC
I
,
II
,
III
−
Aperture diaphragm (iris stop)
The front lens is folded in / out using lever
(3-2/
2
). Turning the turret wheel (3-2/
4
) swings
the brightfield insert and the contrast stops in the
beam path. The abbreviation of the set turret
position (e.g.
D
) is displayed in one of the
windows (3-2/
5
). Slider (3-2/
3
) opens and closes
the aperture diaphragm.
Fig. 3-2
Universal condenser
Summary of Contents for Axioskop 2 FS plus
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...