Then set the required speed step in the range from 0 to 15 via the
Pos. 3
and
Pos. 4
keys The display
value 0 corresponds to the slowest and 15 to the fastest speed step which can be set. The
Pos. 3 key is used to increase the speed step and the Pos. 4 key to reduce it. Each press of the key
changes the setting by one step.
•
Press the
↵
key to confirm the performed setting.
The setting is retained even when the control unit is switched off.
Store and approach the focus position
Up to 4 positions can be stored and approached separately. The positions always refer to the currently
stored zero point. If the zero point is shifted, the stored positions are automatically shifted by the same
amount.
The positions can be set without regard to sequence at any required point.
If a position is stored for the first time, the current position value of this key is simply overwritten.
a)
Store the focus position, e.g. Pos. 2
•
Set the required focus position via the direction keys (or the handwheel).
•
First press the
↵
key and hold it, then press the required position key, e.g. Pos. 2.
•
Release both keys. The set position is now stored on the relevant position key (e.g. Pos. 2).
b)
Approach the focus position, e.g. Pos. 2
•
Press the relevant position key, e.g. Pos. 2. The stored position is then approached automatically.
Set zero point and approach it
Any required position within the travel range of the objective slider can be defined as zero point. The
display is set to zero to allow the measurement of distances. If the zero point is shifted, all the previously
stored positions (Pos. 1 to Pos. 4) are shifted by the same amount.
a)
Set the zero point
•
Use the direction key (or the handwheel) to set the required position for the new zero point.
•
Press the
Set Zero
key to set the current position as the new zero point. The former zero point is
deleted.
b)
Approach the zero point
•
Press the
→←
key to activate the currently stored zero point.
Summary of Contents for Axioskop 2 FS plus
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...