![Zeiss Axioskop 2 FS plus Manual Download Page 41](http://html1.mh-extra.com/html/zeiss/axioskop-2-fs-plus/axioskop-2-fs-plus_manual_3372955041.webp)
Axioskop 2 FS
plus
START-UP
Axioskop 2 FS
MOT
Initial start-up
Carl Zeiss
B 40-076 e 02/01
2-9
2.1.6
Attachment of the condenser
It is recommended to remove the stage carrier
before you insert the condenser.
•
Remove the stage carrier (if not yet done - see
2.1.4) and safely place it on the table.
•
Use lever (2-6/
8
)
to fold out the front lens on
the condenser. Unscrew both centering screws
(2-6/
3
) until their ends are no longer visible.
•
Move condenser carrier (2-6/
2
)
to the lowest
position via drive for height adjustment (2-6/
5
).
•
Insert condenser (2-6/
1
) between condenser
carrier (2-6/
2
) and stage carrier (2-6/
4
). Orient
the locking screw on the underside of the
condenser in the direction of the groove
(2-6/
7
).
•
Press dovetail of condenser against the spring
mount (2-6/
6
) of the condenser carrier until the
condenser can be placed on the condenser
carrier horizontally.
•
Let condenser go smoothly; the screw will lock
in position in the groove (2-6/
7
) at the front.
•
Screw in centering screws until they engage in
the dovetail of the condenser.
Proceed in the same way when attaching the
achromatic condenser 0.8 H D Ph DIC (2-7/
1
). The
item numbers
2
to
7
in Fig. 2-7 are identical to
those in Fig. 2-6.
☞
Attach stage carrier again only after
insertion of the DIC slider.
Fig. 2-6
Attachment of universal condenser
Fig. 2-7
Attachment of achromatic
condenser 0.8 H D Ph DIC
Summary of Contents for Axioskop 2 FS plus
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...