![YSI 2700 SELECT User Manual Download Page 86](http://html.mh-extra.com/html/ysi/2700-select/2700-select_user-manual_915292086.webp)
5-17
4
'
Diagnostic menu functions are used primarily during troubleshooting. In this menu, you will
learn to gain individual control of components in the instrument. For example, you can turn
motors on and off, and even "exercise" motor-controlled devices. You can run pumps to check
fluid movement. You can check probe electrical current and even simulate a calibration or
sample cycle. You can check I/O (input/output) devices such as display, keypad and printer for
proper function. And finally, you can check the status of most sensors in the system. These
include the temperature sensor in the Sample Chamber; the level sensors in the reagent and
waste bottles; the Sipper capacitance sensing; and the optical sensors that control motors, the
Sipper Pump and sense the Test Tube Holder position.
Follow the display prompts and information below to learn about the 2700 SELECT menu
diagnostics mode. As before, you will often need to press [0] to return to a previous menu level.
Occasionally, the instrument will print diagnostic information.
Refer to Figure 5.3 below to become familiar with major components of the instrument.
Additional figures in Sections 3 and 5 may be useful, especially Figure 5.2, Menu Flow Chart.
Figure 5.3
2700 SELECT Major Components
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...