![YSI 2700 SELECT User Manual Download Page 136](http://html.mh-extra.com/html/ysi/2700-select/2700-select_user-manual_915292136.webp)
9-11
When Report Last Transmission (RX) is invoked, the last transmission of the 2700 is sent to
the host. It is used when the response received by the 2700 contains an unknown character, is
not in the correct format, or whenever in doubt due to communications error.
Host to 2700:
RX
Report last transmission
2700 to Host:
[last transmitted record]
When Report Calibration Result (RC) is invoked, the 2700 sends back time, date, chamber
temperature, probe chemistry and probe current from the last calibration cycle. The calibration
cycle may be a user-activated calibration cycle, computer-demanded cycle or an autocalibration
cycle. If no result is found, the 2700 sends back error code "9".
Host to 2700:
RC
Report last calibration result
2700 to Host:
[cal result]
(see format in Figure 9.4)
When Report Sample Result (RS) is invoked, the 2700 sends back the last sample processed.
When combined with an ID number, the 2700 searches its database and sends back the most
recent matching ID result. After the result has been transmitted, its entry in the database is
marked as sent. Another request for the same sample ID result will cause the 2700 to send the
next most recent result for the matching ID result. If no result is found, the 2700 sends back
error code "9".
If the host has received the result in error due to a data transmission error, it should request
retransmission using the RX command.
Host to 2700:
RS
Report last sample result
2700 to Host:
[sam result]
(see format in Figure 9.4)
Host to 2700:
RS< ID>
Report sam ID result
(max = 9 digits)
2700 to Host:
[sam result]
(see format in Figure 9.4)
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...