4-12
This is a direct reading of galactose in solution at the enzyme sensor. The enzyme Galactose
oxidase is immobilized in the enzyme membrane.
Gal O
2
Gal Oxidase
H
2
O
2
+ Galactose Dialdehyde Derivative
System Buffer
YSI 2705
Calibrator Std
n/a *
Linearity Std
n/a *
Black Probe
YSI 2702 Galactose Oxidase Membrane
White Probe
n/a **
Sample Size
25
µ
L
Cal Station
Station #1 Cal Well
Black Probe Parameters
Chemistry
Galactose
Unit of Conc
g/L
Cal value
user's choice (recommend about 2 g/L)
End Point
30 sec
* YSI does not currently offer galactose standard solutions.
** If you have a dual channel unit, either Black or White, or both Black and White probes can
be configured for this chemistry.
Note: See Appendix A if concentration unit conversion is required.
Special Considerations:
»
If sample dilution is required, use reagent water.
»
The sample should be lactose-free. Lactose and other galactosides such as raffinose
and stachyose are substrates for galactose oxidase. They may interfere by producing
artificially high galactose readings.
»
The enzyme membrane integrity test involving the use of YSI 2363 Potassium
Ferrocyanide solution is not informative since, ferrocyanide exists in the 2705 system
buffer.
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...