2-21
SamplingProtocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
Press [2] for Replicates.
Prompt replicate cycle? No
1-No 2-Yes
Since our example setup requests triplicate analysis of the sample, we must activate replicates.
Press [2] for Yes.
Prompt replicate cycle? Yes
1-No 2-Yes
Press [ENTER] to confirm and return to Sampling protocol setup.
SamplingProtocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
Press [3] for ID.
Prompt sample ID? No
1-No 2-Yes
Again, our example setup requests that we use sample identification. Press [2] for Yes.
Prompt sample ID? Yes
1-No 2-Yes
Press [ENTER] to confirm and return to Sampling protocol setup.
SamplingProtocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
Press [4] for Station#.
Enter Sample Station #
> 2
The default station is Station #2 which is where our test tube will be held. Press [ENTER] to
confirm and return to Sampling protocol setup.
SamplingProtocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
Press [0] to return to Run mode setup.
RUN mode setup: 1-SampleProtocol
2-AutoStandby 3-AutoCal 4-Monitor
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...