Users should not open the front access door during operation, internal components may be
susceptible to ESD (electrostatic discharge).
2.
Users are cautioned not to touch the sipper needle during operation, as discussed in the
users manual. The sipper mechanism is equipped with contact-sensing circuitry for the
purposes of fluid detection and human safety. This circuitry may be susceptible to ESD.
ESD to the sipper needle may cause instrument malfunction.
3.
Users should note that RS232 communications may be susceptible to ESD. Data corruption
or serial communications failure may occur. If malfunction occurs, users may restore
proper operation by resetting communications parameters under the “General/RS-232”
setup menu.