1-5
!"#$"
Dextrose (D-Glucose)
YSI Membrane:
2365
Detection Range:
0-9 g/L at 25
µ
L sample size,
0-25 g/L at 10
µ
L sample size
Calibration Point:
2.50 g/L
Linearity Check Point:
9.0 g/L
Precision (CV,n=10):
2%
Linearity (0 to Cal Point):
±
2%
Linearity (Cal to Range Max):
±
5%
Typical Membrane Working Life:
21 days
Membrane O-ring Color:
Red
Ethanol
YSI Membrane:
2786
Detection Range:
0-3.2 g/L
Calibration Point:
2.00 g/L
Linearity Check Point:
3.20 g/L
Precision (CV,n=10):
2%
Linearity (0 to Cal Point):
±
2%
Linearity (Cal to Range Max):
±
5%
Typical Membrane Working Life:
5 days
Membrane O-ring Color:
Green
Potential Substrate Interference:
Methanol
L-Glutamate*
YSI Membrane:
2754
Detection Range:
0-10 mmol/L
Calibration Point:
5.00 mmol/L
Linearity Check Point:
10.0 mmol/L
Precision (CV,n=10):
2%
Linearity (0 to Cal Point):
±
2%
Linearity (Cal to Range Max):
±
5%
Typical Membrane Working Life:
7 days
Membrane O-ring Color:
Yellow
*User may configure instrument to measure as monosodium glutamate (MSG).
L-Lactate (L-Lactic Acid)
YSI Membrane:
2329
Detection Range:
0-2.67 g/L
Calibration Point:
0.50 g/L
Linearity Check Point:
2.67 g/L
Precision (CV,n=10):
2%
Linearity (0 to Cal Point):
±
2%
Linearity (Cal to Range Max):
±
5%
Typical Membrane Working Life:
14 days
Membrane O-ring Color:
Gray
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...