SYMPTOM:
Fail Linearity Test
POSSIBLE CAUSE:
Probe assignment incorrect.
ACTION:
Make correct assignment.
SECTION:
4.0
POSSIBLE CAUSE:
Damaged or old membrane.
ACTION:
Replace membrane.
SECTION:
2.4
POSSIBLE CAUSE:
Contaminated or old calibration or linearity standard.
ACTION:
Repeat test with new standards.
SECTION:
3.4
POSSIBLE CAUSE:
Assigned concentration range beyond practical limits.
ACTION:
Redefine measurement parameters. Remake standards.
SECTION:
4.3
SYMPTOM:
Keypad Disabled
POSSIBLE CAUSE:
Instrument in Remote Control Mode.
ACTION:
Switch communications mode or reset.
SECTION:
9.3
POSSIBLE CAUSE:
Faulty keypad or cable.
ACTION:
Check keypad in I/O diagnostics.
SECTION:
5.4
SYMPTOM:
Printer Does Not Advance
POSSIBLE CAUSE:
Paper or roll jammed.
ACTION:
Open paper well and clear obstruction. Reset instrument
if necessary. If printer still does not advance, turn instrument off
for 30 seconds, then back on.
SECTION:
---
SYMPTOM:
Sipper Does Not Enter Chamber
POSSIBLE CAUSE:
Sipper misaligned.
ACTION:
Check alignment.
SECTION:
2.6
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...