4-14
# # -+!
This is a direct reading of L-glutamate in solution at the enzyme sensor. The enzyme L-Gluta-
mate Oxidase is immobilized in the YSI Glutamate Membrane.
L-Glu O
2
Glut Oxidase
H
2
O
2
+
α
-Ketoglu NH
3
System Buffer
YSI 2357
Calibrator Std
YSI 2755 (5.00 mmol/L glutamate)
Linearity Std
YSI 2756 (10.00 mmol/L glutamate)
Black Probe
YSI 2754 (Glutamate Oxidase Membrane)
White Probe
n/a *
Sample Size
25
µ
L
Cal Station
Station #1 Cal Well
Black Probe Parameters
Chemistry
Glutamate
Unit of Conc
mmol/L
Cal value
5.00
End Point
30 sec
* If you have a dual channel unit, either Black or White, or both Black and White probes can be
configured for this chemistry.
Note: See Appendix A if concentration unit conversion is required.
Special Considerations:
»
If sample dilution is required, use reagent water or 2357 Buffer.
»
If you want to measure MSG (monosodium glutamate, monohydrate) and express the
results as mmol/L of MSG, the calibration parameters listed above are all appropriate.
If you want to express results as a w/v measurement, such as mg/L MSG, then
correction for the sodium and water components is necessary.
To express results as an MSG value, change the glutamate units to a w/v expression,
and enter one of the following: 936 mg/L, 0.936 g/L, or 0.094 %. Use YSI 2755
Calibrator Standard (5.00 mmol/L Glutamate) to calibrate the instrument. Sample
results will then reflect MSG concentration.
Potential Interferences. Tests with equimolar concentrations of other amino acids and related
substances showed little or no interference. On a scale where Glutamate = 100%, those
substances that produced more than 0.5% response were L-Aspartate (0.7), L-Tyrosine (1.4), L-
Histidine (0.7), and L-Glutamine (1.1). Glutamate cannot be ruled out as a contaminant in the
Glutamine solution.
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...