4-17
$
+##
You must have a dual channel version of the 2700 SELECT in order to perform this
measurement. Refer to the sections above on DEXTROSE and L-LACTATE for theoretical and
special considerations. Then follow the instructions below to set up your instrument for
simultaneous determination.
Note: In Section 2 of this manual, the step-by-step procedure for setting up this simultaneous
measurement was described.
System Buffer
YSI 2357
Calibrator Std
YSI 2776 (2.50 g/L dex; 0.50 g/L L-lact)
Linearity Std
YSI 1531 (9.00 g/L dex) and YSI 1530 (2.67 g/L L-lact)
Black Probe
YSI 2329 L-Lactate Membrane
White Probe
YSI 2365 Dextrose Membrane
Sample Size
25
µ
L
Cal Station
Station #1 Cal Well
Black Probe Parameters
Chemistry
L-Lactate
Unit of Conc
g/L
Cal value
0.50
End Point
30 sec
White Probe Parameters
Chemistry
Dextrose
Unit of Conc
g/L
Cal value
2.50
End Point
30 sec
Note: See Appendix A if concentration unit conversion is required.
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...