5-9
Press [2] for MeasParameter. The display will read:
Measurement parameter setup
1-SampleSize 2-CalMethod 3-Black 4-White
“MeasParameter” is an abbreviation for measurement parameters. In this menu, you may select
the sample size and assign the location of the calibrator solution(s). You will also assign specific
chemistry parameters to the Black probe (single channel) or Black and White probes (dual
channel). These parameters include chemistry assignment, calibration value, units of
concentration, and time to endpoint of the enzyme sensor reading.
NOTE: YSI has assigned default values for each of these parameters, but provides flexibility for
the user to change values to optimize for a particular application.
Press [1] for SampleSize. The default setting is 25 microliters. The range is 5 to 65.
Enter sample size in microliter
> 25
Press [ENTER] to confirm and return to Measurement parameter setup.
Measurement parameter setup
1-SampleSize 2-CalMethod 3-Black 4-White
Press [2] for CalMethod. The default setting is One station. The Two Station option is required
for some dual channel analyses, for example, simultaneous dextrose and sucrose determination.
Select calibration method: One station
1-One station 2-Two stations
After pressing [ENTER] to confirm One station, you must specify at which location the
calibration solution will be. The default location is Station #1 (Cal Well). The choices are #1
and/or #2. Number 2 is the Test Tube Holder location. With Accessories such as the YSI 2710
Turntable and YSI 2730 Monitor and Control Accessory, stations #4 and #5 are also options.
Enter Cal Station #
> 1
NOTE: If Two station is selected, you will first be prompted to identify the Black probe cal
station #, and then the White probe cal station #.
Press [ENTER] to confirm and return to Measurement parameter setup.
Measurement parameter setup
1-SampleSize 2-CalMethod 3-Black 4-White
Press [3] for Black. At this level, you will choose chemistry and other black probe measurement
parameters.
Select BLACK chemistry: L-Lactate
0-Backup 1-Next chemistry [ENTER]-Accept
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...