SYMPTOM:
ERROR: Unable to Calibrate (Continued)
POSSIBLE CAUSE:
Failing enzyme membrane.
ACTION:
Perform daily operational checks and replace membrane(s)
if necessary.
SECTION:
3.4
POSSIBLE CAUSE:
Probe surface fouled.
ACTION:
Clean probe surface.
SECTION:
7.5
POSSIBLE CAUSE:
Temperature Probe (auxiliary electrode) fouled.
ACTION:
Clean probe surface.
SECTION:
7.5
SYMPTOM:
ERROR: Undetectable Fluid
POSSIBLE CAUSE:
Low calibrator solution.
ACTION:
Install new calibrator.
SECTION:
2.3
POSSIBLE CAUSE:
Pinched, blocked, leaking or disconnected tube.
ACTION:
Fix or install new tubing.
SECTION:
7.6
POSSIBLE CAUSE:
Calibrator pump not operating properly.
ACTION:
Replace tubing.
SECTION:
7.6
POSSIBLE CAUSE:
Fluid not conductive.
ACTION:
Use saline as diluent.
SECTION:
4.8
POSSIBLE CAUSE:
Low sample level.
ACTION:
Run sample at manual position.
SECTION:
---
SYMPTOM:
Blank Printout
POSSIBLE CAUSE:
Printing on wrong side of paper.
ACTION:
Install paper correctly.
SECTION:
2.6
POSSIBLE CAUSE:
Using wrong type of paper.
ACTION:
Install correct thermal paper (YSI 2751).
SECTION:
2.6
SYMPTOM:
Fail FCN Test
POSSIBLE CAUSE:
Damaged or old membrane.
ACTION:
Replace membrane.
SECTION:
2.4
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...