9-12
Remote Control Commands. The commands outlined above in the "Process" category can be
invoked only in Remote Control Mode. If the system is not in the Remote Control Mode, you
must first invoke the Turn Remote Control On (TR1) command.
Note: For each command from Host to 2700, the 2700 sends back an acknowledgment (A) or
an error code. Error codes may require invoking the Report Status (RY) command to gain
further information on machine status.
Host to 2700:
TR1
Turn on remote control mode
2700 to Host:
A
Acknowledge. 2700 switched to Remote
Control.
2
2700 busy in Run Mode.
9
2700 is in a “halt” condition (page 9-4).
You should invoke this command from Standby or Menu mode. If you attempt to enter Run
Mode by pressing the RUN key locally, then invoke the TR1 command prior to successful
calibration, you may not successfully enter remote control mode.
Alternatively, you may enter Run Mode locally by pressing RUN and waiting for successful
calibration. Then invoke the TR1 command to enter remote control mode.
In a similar manner, you can invoke the Turn Remote Control Off (TR0) command to return
to Result Reporting Mode.
In the Remote Control Mode you may elect to invoke the Turn 2700 Printer Off (TP0)
command. The 2700 responses are outlined below.
Host to 2700:
TP0
Turn 2700 printer off
2700 to Host:
A
Acknowledge. 2700 printer turned off.
1
2700 is not in Remote Control Mode.
In a similar manner, you may use TP1 to turn the 2700 printer on.
Commands to Exit/Enter Run Mode remotely. You may use the commands described below
to enter and exit Run Mode, however, always verify that you are in Remote Control Mode prior
to invoking these commands.
When you invoke TN0 (Exit Run Mode), the instrument exits Run Mode and switches to
Standby Mode or, if there is a “halt condition”, to Menu Mode. Standby Mode may be used to
conserve reagents in certain applications. Menu Mode positions you to remotely recover from
most error conditions.
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...