A-1
In the YSI 2700 SELECT menu selection procedure you have the option to assign units of
concentration. You will find this in the Setup Menu under measurement parameters, black and
white probes. There are default values set based on calibration solutions offered by YSI. Below
is a table of unit conversions for these calibration solutions and other concentrations of interest.
Chemistry
g/L
mg/L
(ppm)
%
(w/v)
mmol/L
mw
(g/mole)
Choline
0.18
175
0.018
1.68
(104)
Dextrose
2.50
2500
0.25
13.89
(180)
Ethanol
2.00
2000
0.20
43.48
(46)
Galactose*
2.50
2500
0.25
13.89
(180)
L-Glutamate
0.73
731
0.07
5.00
(146)
L-Glutamine
0.73
731
0.07
5.00
(146)
L-Lactate
0.50
500
0.05
5.62
(89)
Lactose
5.00
5000
0.50
14.62
(342)
Methanol
1.00
1000
0.10
31.25
(32)
Peroxide*
0.60
600
0.06
17.65
(34)
Sucrose
5.00
5000
0.50
14.62
(342)
* Denotes that YSI does not currently offer this standard value.
If you are preparing your own standard of a value not listed in the preceding table, refer to the
example conversions on the next page to help calculate your unit of choice. If you change units,
remember to enter the appropriate numerical value when prompted by the display. Only the
default values listed above will be automatically converted by the instrument software.
Summary of Contents for 2700 SELECT
Page 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Page 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Page 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Page 49: ...3 10 ...
Page 95: ...5 26 ...
Page 103: ...6 8 ...
Page 117: ...7 14 ...
Page 143: ...9 18 ...
Page 153: ...D 2 ...
Page 155: ...E 2 ...
Page 159: ...F 4 ...
Page 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Page 161: ...G 2 ...
Page 163: ...H 2 ...