4-28
IM 760401-01E
: Displays the result of display A
÷
(display B)
2
When computing impedance (Z), resistance (R), and reactance (X)
Computation example :
V1
A1
SOURCE
LOAD
Var1
A1rms
|X| =
Var1
(A1rms)
2
W1
A1rms
R =
W1
(A1rms)
2
Any
Display A
Display B
Display C
Wiring method
VA1
A1rms
|Z| =
VA1
(A1rms)
2
: Displays the result of (display A)
2
÷
display B
When computing resistance (R)
Computation example :
V1rms
W1
R =
(V1rms)
2
W1
Any
Display A
Display B
Display C
Wiring method
V1
A1
SOURCE
LOAD
4.11 Performing Four Arithmetical Operation
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...