4-27
IM 760401-01E
Setting Measurement Conditions and Measurement Range
4
Computation example 2:
3P3W
Display A
Display B
Display C
Wiring method
Σ
W (= W1 + W3) W2
Σ
W – W2
W2
W1
W3
Inverter
Computation example 3:
3P3W
Display A
Display B
Display C
Wiring method
W2
Σ
W (= W1 + W3) W2 –
Σ
W
W2
W1
W3
Inverter
: Displays the result of display A
×
display B.
Useful when setting a function other than VA (apparent power) for display A and
displaying VA on display C.
Computation example :
Any
Display A
Display B
Display C
Wiring method
V1rms
A1rms
A1rms
×
A1rms
: Displays the result of display A
÷
display B.
When computing the absolute value of impedance
Computation example :
Any
Display A
Display B
Display C
Wiring method
V1rms
A1rms
|Z| =
V1rms
A1rms
V1
A1
SOURCE
LOAD
: Displays the result of display A
÷
display B.
Line voltage ratio and line current ratio for a three-phase wiring can be determined.
Computation example :
A1rms
A3rms
A1rms
A3rms
3P3W
Display A
Display B
Display C
Wiring method
V1rms
V3rms
V1rms
V3rms
V1
SOURCE
LOAD
V3
A1
A3
4.11 Performing Four Arithmetical Operation
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...