14-31
IM 760401-01E
Communication Commands 2 (System of Commands Complying to the IEEE 488.2-1992 Standard)
14
MEASure[:NORMal]:ITEM:PRESet
Function
Sets the ON/OFF pattern for all communication
outputs of the normal measurement function.
Syntax
MEASure[:NORMal]:ITEM:PRESet {NORMal|
INTEGrate|CLEar}
Example
MEASURE:NORMAL:ITEM:PRESET NORMAL
Description The following three patterns can be selected.
The same setting applies to the current all
elements or
∑
.
NORMal
: V/A/W -> ON, others -> OFF
INTEGrate
: W/WH/AH/TIME -> ON, others ->
OFF
CLEar
: all items -> OFF
MEASure[:NORMal]:ITEM:{TIME|MATH}
Function
Sets the communication output of
{elapsed
integration time|MATH}
ON/OFF, queries
about the current setting.
Syntax
MEASure[:NORMal]:ITEM:{TIME|
MATH} {<Boolean>}
MEASure[:NORMal]:ITEM:{TIME|MATH}?
Example
MEASURE:NORMAL:ITEM:TIME OFF
MEASURE:NORMAL:ITEM:TIME?
-> :MEASURE:NORMAL:ITEM:TIME 0
MEASure[:NORMal]:ITEM:
<normal measurement function>?
Function
Queries communication output settings of the
normal measurement function.
Syntax
MEASure[:NORMal]:ITEM:
<normal measurement function>?
<normal measurement function>={V|A|W|
VA|VAR|PF|DEGRee|VHZ|AHZ|WH|WHP|WHM|
AH|AHP|AHM|VPK|APK}
Example
MEASURE:NORMAL:ITEM:V?
-> :MEASURE:NORMAL:ITEM:V:ELEMENT1 1;
ELEMENT2 1;ELEMENT3 1;SIGMA 1
Description For the meanings of the symbols of functions,
see Note on page 14-14.
MEASure[:NORMal]:ITEM:
<normal measurement function>[:ALL]
Function
Sets the communication output concerning all
elements or
∑
ON/OFF at once.
Syntax
MEASure[:NORMal]:ITEM:<normal
measurement function>[:ALL] {<
Boolean>}
Example
MEASURE:NORMAL:ITEM:V:ALL ON
MEASure[:NORMal]:ITEM:
<normal measurement function>:ELEMent<x>
Function
Sets the communication output concerning each
element ON/OFF, queries the current setting.
Syntax
MEASure[:NORMal]:ITEM:
<normal measurement function>:
ELEMent<x> {<Boolean>}
MEASure[:NORMal]:ITEM:
<normal measurement function>:
ELEMent<x>?
14.3 Commands
<x>= 1
(WT210 single-phase
model)
1, 3
(WT230 three-phase
three-wire model)
1 to 3 (WT230 three phase four-
wire model)
Example
MEASURE:NORMAL:ITEM:V:ELEMENT1 ON
MEASURE:NORMAL:ITEM:V:ELEMENT?
-> :MEASURE:NORMAL:ITEM:V:
ELEMENT1 1
MEASure[:NORMal]:ITEM:
<normal measurement function>:SIGMa
Function
Sets the communication output concerning
∑
ON/OFF, queries the current setting.
Syntax
MEASure[:NORMal]:ITEM:
<normal measurement function>:
SIGMa {<Boolean>}
MEASure[:NORMal]:ITEM:
<normal measurement function>:SIGMa?
Example
MEASURE:NORMAL:ITEM:V:SIGMA ON
MEASURE:NORMAL:ITEM:V:SIGMA?
-> :MEASURE:NORMAL:ITEM:V:SIGMA 1
MEASure[:NORMal]:VALue?
Function
Queries normal measured/computed data set by
commands other than
“
MEASure[:NORMal]:ITEM
” (ASCII format).
Syntax
MEASure[:NORMal]:VALue?
Example
MEASURE:NORMAL:VALUE?
-> 10.04E+00,10.02E+00,10.03E+00,
49.41E+00,...
Description • The renewal of normal measured/computed
data output here occures when bit0 (UPD) of
the condition register (refer to page 14-47)
changes from high to low. For more details,
refer to 14.2.6.
• For the output format of normal measured/
computed data, refer to page 14-32.
• When the harmonic measurement function is
ON, harmonic measurement data will be
returned (same as the response to
MEASure:HARMonics:VALue?
).
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...