14-62
IM 760401-01E
Errors in communications execution (200 to 299)
Code
Message
Action
Reference
221
Setting conflict
Check the relevant setting.
14.3
222
Data out of range
Check the setting range.
14.3
223
Too much data
Check the data byte length.
14.3
224
Illegal parameter value
Check the setting range.
14.3
241
Hardware missing
Check availability of options.
—
260
Expression error
Equation is not allowed.
—
270
Macro error
Does not conform to the macro definition specified in IEEE488.2. —
272
Macro execution error
Does not conform to the macro definition specified in IEEE488.2. —
273
Illegal macro label
Does not conform to the macro definition specified in IEEE488.2. —
275
Macro definition too long
Does not conform to the macro definition specified in IEEE488.2. —
276
Macro recursion error
Does not conform to the macro function specified in IEEE488.2.
—
277
Macro redefinition not allowed Does not conform to the macro definition specified in IEEE488.2. —
278
Macro header not found
Does not conform to the macro definition specified in IEEE488.2. —
Error in communication Query(400 to 499)
Code
Message
Action
Reference
410
Query INTERRUPTED
Check transmission/reception order.
14-4
420
Query UNTERMINATED
Check transmission/reception order.
14-4
430
Query DEADLOCKED
Limit the length of the program message including <PMT> to
14-5
1024 bytes or less.
440
Query UNTERMINATED after
Do not enter any query after *IDN? and *OPT?.
—
indefinite response
Errors in Execution (800 to 899)
Code
Message
Action
Reference
813 to 819 Invalid operation
Refer to 15.4
15-14
830 to 833 Internal memory access error
Refer to 15.4
15-14
841 to 847 Integrator execute error
Refer to 15.4
15-14
Error in System Operation (912)
Code
Message
Action
Reference
912
Fatal error in Communication
Service is required.
—
driver
Warnings (350, 390)
Code
Message
Action
Reference
350
Queue overflow
Read out the queue.
14-48
390
Overrun error
Adjust the baud rate.
11-5
(only for RS-232-C)
Note
The warning code 350 only appears in case of an overflow of the error queue. The error
which occurs in case of clearing the STATus:ERRor? will not appear on the screen.
14.11 Communication-related Error Messages
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...