14-35
IM 760401-01E
Communication Commands 2 (System of Commands Complying to the IEEE 488.2-1992 Standard)
14
Data Format/Output Format of Normal Measurement and Harmonic Measurement Data
<BINARY>
The data format and output format of normal measurement and harmonic measurement
data in binary format that are output by the
MEASure[:NORMal]:BINary?
or
MEASure:HARMonics:BINary?
command are as follows:
Data Format
Measured data consists of a 2-byte header and 4-byte data (total of 6 bytes).
Header (2 bytes)
Data (4 bytes)
Header structure
The header consists of a 1-byte status and 1-byte ASCII conversion information.
Status (1 byte)
ASCII conversion information (1 byte)
• Status byte
b7
b6
b5
b4
b3
b2
b1
b0
• b7 to b4: Lead/Lag phase information
(This information is meaningful only when the data is a phase angle (DEGR).
For all other cases, the value is always 0000.)
0000: Lag
0001: Lead
0010: Cannot be detected
• b3 to b0: Data status
0000: Normal
0001: Overrange (--oL-)
0010: Voltage peak over
0011: Current peak over
0100: Power factor error (PFErr)0101: Phase angle error (dEGEr)
0110: Frequency error (ErrLo) 0111: Frequency error (ErrHi)
1000: Computation overflow (--oF-)
1001: PLL error (FreqEr)
1010: No data (-----)
• ASCII conversion information
b7
b6
b5
b4
b3
b2
b1
b0
• b7 to b4: Prefix unit
0000: None (E+0)
0001: m (E–3)
0010: k (E+3)
0011: M (E+6)
• b3 to b0: Decimal point position
0000: _ _ _ _ _ .
0001: _ _ _ _ . _
0010: _ _ _ . _ _
0011: _ _ . _ _ _
0100: _ . _ _ _ _
0101: . _ _ _ _ _
Data structure
Data is in 4-byte IEEE single-precision floating point format.
S E E E E E E E E M M M M M M M M M M M M M M M M M M M M M M M
The byte order is MSB first.
Physical value = (–1)
S
×
(2(E
–127
))
×
(1+M/(2
23
))
S: Sign bit, 0 or 1
E: Exponent in the range of 0 to 254
M: Mantissa as a 23-bit binary value
• For elapsed integration time
Outputs the time in units of seconds. For example, 0x45610000(=3600) if the time is 1
hour (1:00:00).
• For overrange and computation overflow
0x7E94F56A (=9.9E+37)
• For no data
0x7E951BEE (=9.91E+37)
Output format
The output format (output order of measured data) is the same as the ASCII format.
14.3 Commands
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...