10-3
IM 760401-01E
GP-IB Interface
(Option)
10
10.2 Responses to Interface Messages, and
Remote/Local Modes
Responses to Interface Messages
IFC (Interface Clear)
Unaddresses talker and listener.
REN (Remote Enable)
Transfers the instrument from local control to remote control.
GTL (Go To Local)
Transfers the instrument from remote control to local control.
SDC (Selective Device Clear), DCL (Device Clear)
Cleasrs GP-IB input/output buffer, and resets an error. The setup information and
measurement state are not affected. DCL is applicable to all devices on the bus,
whilst DSC is applicable to designated devices only.
GET (Group Execute Trigger)
Updates the measured/computed data when in hold mode.
Same function as the SHIFT + HOLD (TRIG) key.
LLO (Local Lockout)
Invalidates the LOCAL key on the front panel to inhibit transfer from remote control to
local control.
Switching between Remote and Local Mode
When switched from local to remote mode
The REMOTE indicator will light up, and all panel keys except the LOCAL key cannot
be operated. Setup parameres entered in the local mode will be retained.
When switched from remote to local mode
The REMOTE indicator will extinguish and all panel keys can be operated. Setup
parameters entered in the remote mode will be retained.
Valid keys for remote control
Pressing the LOCAL key in remote control will switch the instrument to local control.
However, this is not possible in case the Local Lockout has been set by the controller.
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...