Communication Commands 2 (System of Commands Complying to the IEEE 488.2-1992 Standard)
14
14.3.12 SAMPle Group
The commands in the SAMPle group are used to make settings relating to sampling. The commands allow you to
make the same settings and inquiries as when the [HOLD] key on the front panel or the “u.rAtE” menu of the [SETUP]
key is used.
:SAMPle
:
HOLD
<Space>
OFF
ON
<NRf>
?
RATE
<Space>
<time>
?
?
;
14.3 Commands
SAMPle?
Function
Queries all settings related to sampling.
Syntax
SAMPle?
Example
SAMPLE? -> :SAMPLE:HOLD 0
SAMPle:HOLD
Function
Sets to hold the output of data (display,
communication)/queries the current setting.
Syntax
SAMPle:HOLD {<Boolean>}
SAMPle:HOLD?
Example
SAMPLE:HOLD ON
SAMPLE:HOLD? -> :SAMPLE:HOLD 1
SAMPle:RATE
Function
Sets the display update rate or queries the
current setting.
Syntax
SAMPle:RATE {<time>}
<time> = 0.1 to 5 s (0.1, 0.25, 0.5,
1, 2, 5)
Example
SAMPLE:RATE 0.25S
SAMPLE:RATE?
-> :SAMPLE:RATE 0.25E+00