Setting Measurement Conditions and Measurement Range
4
Chapter 4 Setting Measurement Conditions and Measurement Range
4.1
Selecting the Measurement Mode
Keys
CHECK RANGE
MODE
UPDATE
SCALING
AVG
LINE
FREQ
STORE
RECALL
HARMONICS KEY LOCK
FILTER
A
B
C
VOLTAGE
VOLTAGE
VOLTAGE
AUTO
MODE
CURRENT
AUTO
RANGE
MAX HOLD
TRIG
CAL
INTEGRATOR
INTEG SET
SHIFT
KEY LOCK
1P 3W
3P 3W
3P 4W
3V 3A
OUTPUT
HARMONICS
MEMORY
REMOTE
MEAN
CURRENT
RMS
DC
FUNCTION
ELEMENT
MAX HOLD
HOLD
ENTER
RESET
WIRING
START
STOP
LOCAL
SETUP
1
m
V
k
A
V
TIME
ar
deg
h
h
VA
M W
m
V
PF
k
A
M W
m
V
Hz
k
A
M W
2
3
FUNCTION
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
%
The explanation given in this section uses WT230 as an example. For the differences between
the WT210 and the WT230, see section 2.2, “Operation Keys and Functions/Element Display.”
Procedure
RMS
VOLTAGE
MEAN
DC
MODE
SHIFT
MODE
VOLTAGE
VOLTAGE
VOLTAGE
SHIFT
MODE
SHIFT
Explanation
Measurement Mode
One of the following measurement modes can be selected for measurement of voltage
and current. The initial value is “RMS”.
Indicator
Voltage
Current
RMS
Measures and displays true
Measures and displays true RMS
RMS value
value
VOLTAGE MEAN
Displays rectified mean value
Measures and displays
calibrated to the RMS value
true RMS value
DC
Displays DC value obtained by
Displays DC value obtained by averaging
averaging the input signal
the input signal
Theoretical Equations
• RMS
This mode is selected to display input voltage or current as a true RMS value.
T
1
T
0
(t)
2
dt
f (t): input signal
T:
one period of the input signal
• VOLTAGE MEAN
This mode is selected to display input voltage or current as a rectified mean value
calibrated to the RMS value. Since a sine wave is used for calibration, the value
displayed will be the same as that obtained in RMS mode if a sine wave is
measured. The value displayed will be different from that obtained in RMS mode if
a distorted or DC waveform is measured.
T
1
(t)
dt
2 2
·
π
T
0
f (t): input signal
T:
one period of the input signal
• DC
This mode is selected when the input voltage or current is DC. The input signal is
averaged and the result is displayed.
Summary of Contents for WT210
Page 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Page 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Page 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Page 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Page 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...