Appendix 2 Application Examples
Source and Measurement
1.
Use the first GS610 to apply a constant voltage V
CE
across the collector and the
emitter of the transistor.
2.
Use the second GS610 to supply current pattern I
B
(base current) that you
created in step 1 of “Preparation” across the base and the emitter of the
transistor. The first GS610 measures collector current I
C
by synchronizing to the
current pattern using the external trigger signal.
3.
The measurement result data is created in the internal memory (GS610RAM) of
the first GS610.
4.
Transfer (Copy) the measurement result file in the internal memory
(GS610RAM) of the first GS610 and the pattern file saved in the internal
memory (GS610ROM) of the second GS610 to your PC via the USB.
5.
Using a general-purpose spreadsheet application on your PC, save the sweep
pattern and the measured results to a single file.