18.1 Source Section
DC Voltage Source
Range Setting Source Range Resolution Max. Load Accuracy (One Year)
Temperature Coefficient
Current
±
(% of s V + V
×
Io/
±
(% of s V + V
×
Io/
f.s of the limiter range)
f.s of the limiter range)/
°
C
200 mV
±
205.000 mV
1
µ
V
±
3.2 A
0.02+200
µ
V+80
µ
V (400
µ
V) 0.002+20
µ
V+8
µ
V (40
µ
V)
2 V
±
2.05000 V
10
µ
V
±
3.2 A
0.02+300
µ
V+100
µ
V (500
µ
V) 0.002+30
µ
V+10
µ
V (50
µ
V)
12 V
±
12.0000 V
100
µ
V
±
3.2 A
0.02+2 mV+800
µ
V (3 mV)
0.002+200
µ
V+80
µ
V (300
µ
V)
20 V
±
20.5000 V
100
µ
V
±
2 A
0.02+2 mV+800
µ
V (5 mV)
0.002+200
µ
V+80
µ
V (500
µ
V)
30 V
±
30.000 V
1 mV
±
2 A
0.02+20 mV+5 mV (30 mV)
0.002+2 mV+500
µ
V (3 mV)
60 V
±
60.000 V
1 mV
±
1 A
0.02+20 mV+6 mV (40 mV)
0.002+2 mV+600
µ
V (4 mV)
110 V
±
110.000 V
1 mV
±
0.5 A
0.02+20 mV+8 mV (70 mV)
0.002+2 mV+800
µ
V (7 mV)
Accuracy: One year accuracy at 23
±
5
°
C.
Temperature coefficient: Add the temperature coefficient at 5 to 18
°
C and 28 to 40
°
C.
The values inside the parentheses are those when the limiter range is 3 A.
DC Current Source
Range Source
Resolution Max. Load Accuracy (One Year) Temperature
Setting Range
Voltage
±
(% of A)
Coefficient
±
(% of A)/
°
C
20
µ
A
±
20.5000
µ
A
100 pA
±
110 V
0.03+50 nA
0.003+5 nA
200
µ
A
±
205.000
µ
A
1 nA
±
110 V
0.03+300 nA
0.003+30 nA
2 mA
±
2.05000 mA 10 nA
±
110 V
0.03+3
µ
A
0.003+300 nA
20 mA
±
20.5000 mA 100 nA
±
110 V
0.03+30
µ
A
0.003+3
µ
A
200 mA
±
205.000 mA 1
µ
A
±
110 V
0.03+300
µ
A
0.003+30
µ
A
0.5 A
±
0.50000 A
10
µ
A
±
110 V
0.03+5 mA
0.003+500
µ
A
1 A
±
1.00000 A
10
µ
A
±
60 V
0.03+5 mA
0.003+500
µ
A
2 A
±
2.00000 A
10
µ
A
±
30 V
0.03+5 mA
0.003+500
µ
A
3 A
±
3.20000 A
10
µ
A
±
12 V
0.03+5 mA
0.003+500
µ
A
Accuracy: One year accuracy at 23
±
5
°
C.
Temperature coefficient: Add the temperature coefficient at 5 to 18
°
C and 28 to 40
°
C.
Current Limiter
|Limiter Value|
1
Range Setting
Resolution
Minimum Limit Value
2
0.10
µ
A to 20.00
µ
A
20
µ
A
10 nA
100 nA
20.1
µ
A to 200.0
µ
A
200
µ
A
100 nA
1
µ
A
0.201 mA to 2.000 mA
2 mA
1
µ
A
10
µ
A
2.01 mA to 20.00 mA
20 mA
10
µ
A
100
µ
A
20.1 mA to 200.0 mA
200 mA
100
µ
A
1 mA
0.201 A to 3.200 A
3 A
1 mA
10 mA
1 When tracking is OFF, |Hi| or |Lo|, whichever is greater.
2 Minimum value when tracking is OFF.
Voltage Limiter
|Limiter Value|
3
Range Setting
Resolution
Minimum Limit Value
4
1.0 mV to 200.0 mV
200 mV
100
µ
V
1 mV
0.201 V to 2.000 V
2 V
1 mV
10 mV
2.01 V to 20.00 V
20 V
10 mV
100 mV
20.1 V to 110.0 V
110 V
100 mV
1 V
3 When tracking is OFF, |Hi| or |Lo|, whichever is greater.
4 Minimum value when tracking is OFF.
Chapter 18 Specifications