5-16
IM 765501-01E
5.9
Offset Calibration
Procedure
MEASURE
STORE
AUTO
RANGE
MENU
MATH
NULL
COMPARE
MEASURE
VALUE
LOCAL
POWER
MISC
TIME
KEY LOCK
SHIFT
V I
MODE
AUTO
RANGE
LIMIT
VS IS
MENU
SOURCE
OUTPUT CONTROL
SWEEP
START
ZERO
SOURCE
TRIG
MODE
OUTPUT
RANGE
Ω
RECALL
DISPLAY
ESC
• To exit the menu during operation, press
ESC
at the upper left of the operation
panel.
• In the procedural explanation below, the
phrase “
rotary knob, numeric keys, <,
and >
” are used to refer to the operation
of selecting or setting items or entering
values using the
rotary knob
,
numeric
keys
, , and keys (BS key and
right arrow key). For details on the
operation using the rotary knob, numeric
keys, , and , see sections 3.8 or
3.9.
1.
Press
SHIFT+VS/IS(SOURCE)
to display the SOURCE menu.
2.
Press the
CalExec
soft key to carry out the calibration.
Explanation
Carry out calibration when correcting the offset drift of the source value such as due to
temperature changes.
Note
The offset calibration is valid only for the range setting at which the calibration is carried out.
The calibration setting is initialized when the power is cycled.
<<Corresponding Command Mnemonic>>
*CAL?